Watch tutorials, webinars and informative videos about PreSens optical sensor systems.
Bare Fiber Oxygen Microsensor for High Resolution Trace Oxygen Measurements
Implantable Oxygen Microsensor IMP-PSt8
This tiny probe has a tip size of 250 ?m while the outer diameter is 1000 ?m. The microsensor tip is not mounted into any additional housing, so it can be applied in various customized applications. The bare plastic fiber tip can be mounted to your own housings, steel tubes and micro respirometer chambers etc. The small outer diameter even allows insertion into implantable Venflon-tubes. Moreover the microsensor can be deployed in sealed containers to measure the oxygen content directly. The IMP-PSt8 has a measurement range of 0 – 10 % oxygen, with a limit of detection of 0.007 % O2.
- High spatial resolution (250 ?m)
- High temporal resolution (t90 < 3 sec.)
- Integration into customized housings or set-ups
- No consumption of oxygen
- Measurement in headspace & liquids
Technical
| Specifications | Gaseous & Dissolved O2 | Dissolved O2 |
|---|---|---|
| *at + 20 °C, after two-point calibration **TF = flat sensor tip with 250 ?m diameter | ||
| Measurement range | 0 – 10 % O2 0 – 100 hPa | 0 – 4.5 mg/L 0 – 140 ?mol/L |
| Limit of detection | 0.007 % oxygen | 3 ppb |
| Resolution | ± 0.002 % O2 at 0.008 % O2 | ± 0.7 ppb at 3 ppb |
| Accuracy at + 20 °C* | ± 3 ppb or ± 3 % rel. | |
| Measurement temperature range | from 0 to + 50 °C | |
| Response time (t90) | TF**: < 3 sec. (gas) | TF**: < 10 sec. (liquid) |
| Properties | ||
| Compatibility | Aqueous solutions | |
| No cross-sensitivity | pH 1 – 14 | |
| Cross-sensitivity | Organic solvents, such as acetone, toluene, chloroform or methylene chloride | |
| Disinfection procedure | Disinfection with EtOH (70 % v/v) or isopropanol (100 % v/v) possible | |
| Cleaning procedure | Ethanol, soap solution, ddw | |
| Calibration | Two-point calibration with oxygen-free environment (nitrogen, sodium sulfite) and a second calibration value optimally between 2 % and 4 % oxygen | |
Related products
Resources
FAQs
Manuals
Oxygen Microsensors NTH / IMP / PM TF-250 (plastic fiber)
Oxygen Microsensors NTH / IMP / PM / FTCM (glass fiber)
Getting Started: Microx 4 & Microx 4 trace
PreSens Oxygen Calculator
Microx 4 & Microx 4 trace
Manual Micromanipulator MM
Automated Micromanipulator AM
User Guide: Automated Micromanipulator
Brochures
Software
PreSens Oxygen Calculator version 3.1.1 - Windows 7/8/10